Back to articles
General Microbiology12 min read

Sterilization and Disinfection Methods

The framework hospitals actually use to decide how sterile a device needs to be, sterilization vs. disinfection vs. decontamination explained clearly, the Spaulding Classification, and the full mechanism of every method linked below.

The device that was cleaned, just not enough

Endoscopes present a genuinely difficult reprocessing problem: they're expensive, reusable, contain long narrow internal channels that are hard to fully access, and they contact mucous membranes deep inside the body during almost every procedure. In the mid-2010s, a series of real, well-documented U.S. healthcare-associated infection investigations traced clusters of carbapenem-resistant Enterobacteriaceae (CRE) infections back to inadequately reprocessed duodenoscopes, despite staff having followed a cleaning protocol.

The failure wasn't carelessness. It was that a device contacting mucous membranes needs to reach a specific, defined level of microbial killing, high-level disinfection at minimum, and the combination of the device's complex internal geometry and the cleaning protocol in use didn't reliably achieve that level inside every channel, every time. A blood pressure cuff, which only ever touches intact skin, can be safely reused after a much lower level of cleaning. An endoscope cannot. Treating the two as if they belonged in the same category, even informally, is exactly the kind of gap that leads to outbreaks like this one.

The framework that exists specifically to prevent this confusion, deciding exactly how sterile a given device needs to be, based on what part of the body it will actually touch, is called the Spaulding Classification, and it's the organizing idea behind everything in this article.

Sterilization is the process of killing or removing all microorganisms, including highly resistant bacterial endospores. Depending on the material, sterilization can be achieved by moist heat (autoclave), dry heat (hot air oven), ethylene oxide gas, or filtration.

However, sterilization is not always attainable or practical, for example, in fresh produce processing. In those cases, rapid microbial growth is limited instead by decontamination and disinfection.

Methods of sterilization and disinfection - Methods of sterilization and disinfectionFigure: Methods of sterilization and disinfection

Disinfection is the killing of pathogenic organisms, but not necessarily all microorganisms. Pathogenic bacteria are destroyed during disinfection, but some microorganisms, including bacterial endospores, may remain viable. Disinfectant choice depends on the intended use, inanimate objects versus skin surfaces. Common disinfectants include phenol-containing compounds, quaternary ammonium compounds, alcohol, chlorine, and iodine.

Decontamination is the removal of pathogenic microorganisms from inanimate objects or surfaces so that items are safe to handle or dispose of, the least rigorous of the three terms. Simply wiping a table after a meal, for example, removes contaminating microorganisms and their potential nutrients, without claiming to kill anything specific.

The Spaulding Classification

Deciding which of these three processes a given item actually needs isn't guesswork. The Spaulding Classification, developed by Earle Spaulding and adopted by the CDC, sorts medical devices into three categories based on the infection risk of how they're used, not on how the device looks or feels:

Category Definition Required processing
Critical Enters normally sterile tissue, the vascular system, or blood flow Sterilization (all microbial life, including endospores, must be destroyed) — e.g., surgical instruments, cardiac catheters, implants, needles
Semicritical Contacts mucous membranes or non-intact skin, without penetrating sterile tissue High-level disinfection at minimum — e.g., endoscopes, respiratory therapy equipment
Noncritical Contacts only intact skin Low- to intermediate-level disinfection — e.g., blood pressure cuffs, stethoscopes

This is exactly the framework the duodenoscope outbreaks above ran into: a semicritical device needs high-level disinfection every single time, in every channel, and any gap between that requirement and what actually happens during reprocessing becomes a real infection risk.

Factors Affecting Sterilization and Disinfection Success

Several factors determine whether a given sterilization or disinfection method actually succeeds: the microbial load and amount of organic material present, the type of organism involved (bacterial spores are far more resistant than most vegetative bacteria or enveloped viruses), the concentration and exposure time of the germicide, the physical nature of the surface, and temperature and pH.

For the full breakdown of each factor, plus the complete hierarchy of microbial resistance, see Disinfection: Methods and Uses.

Sterilization Methods

Physical Agents of Sterilization

Heat, radiation, and filtration are the most commonly used physical sterilization methods, with heat being the most widely used overall.

Incineration. Treating infectious waste by burning it to ash at 870–980°C. Incineration is the most complete method available for destroying pathogenic microorganisms, bacterial endospores, and even prions, though toxic air emissions and heavy metals in the resulting ash have limited its broader use.

Moist heat sterilization. Using autoclaves, moist heat (saturated steam under pressure) sterilizes bio-hazardous waste and heat-stable objects, the fastest and simplest physical sterilization method available. The two common sterilization parameters are 121°C for 15 minutes and 132°C for 5 minutes. Read the full mechanism, including boiling, tyndallization, and pasteurization, in Moist Heat Sterilization, or the full autoclave procedure and parts in Autoclave Sterilization.

Medical autoclave - Medical Autoclave SterilizerImage: Medical Autoclave

Dry heat sterilization. Requires longer exposure (1.5–3 hours) and higher temperatures (160–180°C) than moist heat, used for glassware, oils, petrolatum, and powders. Read more in Dry-Heat Sterilization, or about the equipment itself in Hot Air Oven: Parts, Types, and Uses.

hot air oven - Hot air ovenSource: AcumenlabwareSource: Autoclave

Radiation. Ionizing radiation generates reactive species (electrons, hydroxyl and hydride radicals) that damage DNA, lipids, and protein, killing the irradiated cell. The FDA has approved radiation for sterilizing surgical supplies, disposable labware, drugs, and tissue grafts; the WHO has approved it for decontaminating foods susceptible to microbial contamination, such as fresh produce, meat, poultry, and spices. Read the full mechanism, including non-ionizing UV disinfection, in Radiation Sterilization.

Filtration. Passing an aqueous liquid through the tiny pores of a membrane filter traps microorganisms present in the liquid, rendering it sterile, without killing anything. High-efficiency particulate air (HEPA) filters serve the equivalent role for air filtration.

Concept of filtration - Filtration sterilization by removing organismsRead the full mechanism, including what filtration doesn't remove, in Filtration Sterilization.

Chemical Agents of Sterilization

Chemicals used to destroy all forms of life, including spores, are called chemical sterilants or biocides. Ethylene oxide and hydrogen peroxide are common examples. The same chemicals, used at lower concentration or for shorter contact time, function as disinfectants instead. Ethylene oxide (EtO) is the most common chemical sterilant for heat-sensitive objects; vapor-phase hydrogen peroxide is used to sterilize HEPA filters in biological safety cabinets and both metal and nonmetal medical instruments.

Disinfection Methods

Physical Methods of Disinfection

Boiling and pasteurization. Boiling at 100°C kills most vegetative bacteria and non-sporing organisms but does not reliably kill bacterial endospores, making it a disinfection method, not a sterilization one. Pasteurization uses controlled heat to reduce pathogens (including those causing tuberculosis, brucellosis, and typhoid) without sterilizing the product.

For the full comparison of boiling, tyndallization, pasteurization, and true steam sterilization, see Moist Heat Sterilization.

Non-ionizing radiation (UV). UV rays are long-wavelength, low-energy radiation that can damage microbial DNA directly, but they don't penetrate solid, opaque, or light-absorbing surfaces. This is why UV disinfection is limited to directly exposed surfaces, such as the work surface of a biological safety cabinet after use.

UV light Radiation Sterilization - Use of UV light for sterilizationImage: Use of UV Light for Sterilization

For the full mechanism, including how UV was used during the COVID-19 pandemic, see Radiation Sterilization.

Chemical Methods of Disinfection and Levels of Disinfection

Chemical disinfectants are classified by how much they can kill: low-level disinfectants handle most vegetative bacteria and some fungi/viruses; intermediate-level disinfectants add mycobacteria; high-level disinfectants kill everything except large numbers of bacterial spores. Common chemical disinfectants include alcohols, phenolics, aldehydes, halogens (chlorine and chlorine compounds), hydrogen peroxide, peracetic acid, and quaternary ammonium compounds. Bleach (sodium hypochlorite) is commonly used to disinfect food-preparation surfaces.

Chemical disinfectants - Chemicals disinfectantsImage source: Chemical disinfectants

For the full activity-level table, the hierarchy of microbial resistance, and why some disinfectants can't be relied on against spore-forming pathogens, see Disinfection: Methods and Uses.

Antiseptics

Antiseptics are disinfectants formulated for use on living tissue, such as skin or mucous membranes, rather than inanimate objects. They matter for personal laboratory safety and for preparing patients before invasive procedures such as blood draws or surgery.

Iodine compounds (tincture-iodine or povidone-iodine) are widely used antiseptics. A combination of 70% ethyl alcohol followed by an iodophor is the most common approach to skin disinfection before drawing blood cultures or performing surgery.

Clinical Use

Sterilization and disinfection are basic components of hospital infection control. According to the CDC, more than 99,000 deaths occur each year in the United States alone due to hospital-acquired infections, making the correct application of the framework above a matter of real clinical consequence, not just an exam topic.

Devices intended to enter normally sterile sites (surgical instruments, cardiac catheters, implants, needles) should be sterilized by steam, plasma, or ethylene oxide, per the Critical category above. Devices contacting only intact skin (blood pressure cuffs, stethoscopes) are decontaminated with an intermediate- or low-level disinfectant, per the Noncritical category.

Commonly used disinfection or sterilization methods and their clinical use:

Clinical Use Commonly Used Disinfection or Sterilization Method
Surgical hand disinfection Chlorhexidine
Surgical site skin preparation Iodophor
Skin preparation for venipuncture or immunization 70% ethanol
Disinfect skin before blood culture or vascular catheter insertion Tincture of iodine followed by 70% ethanol, or iodophor, or chlorhexidine
Cleanse wounds Thimerosal, chlorhexidine, hydrogen peroxide
Cleanse burn wounds Silver sulfadiazine
Clean up blood spill from a hepatitis B/C patient Hypochlorite (bleach)
Sterilize heat-sensitive materials (endoscopes, respiratory equipment) Ethylene oxide or glutaraldehyde
Sterilize heat-resistant materials (surgical gowns, drapes) Autoclave
Sterilize intravenous solutions Filtration
Disinfect operating theatre air (when not in use) Ultraviolet light
Disinfect operating room floor Benzalkonium chloride (Lysol)
Disinfect stethoscope 70% ethanol or isopropyl alcohol
Preserve vaccines Thimerosal

How to Remember

The "risk elevator" analogy for the Spaulding Classification. Picture medical devices riding an elevator with three floors, based on how deep into the body they go. Ground floor: intact skin only (noncritical), a quick clean is enough. Middle floor: mucous membranes (semicritical), the device needs high-level disinfection before it's allowed back on. Top floor: sterile tissue or bloodstream (critical), nothing short of full sterilization is acceptable. The higher the floor, the less room there is for anything to survive.

Anchor for sterilization vs. disinfection vs. decontamination: sterilization means zero survivors, no exceptions, spores included. Disinfection means the dangerous ones are dead, but spores might still be there. Decontamination is the lowest bar of all: just "safe enough to handle," with no specific claim about what's still alive.

Anchor for the hook: the duodenoscope outbreaks happened because a semicritical device needs high-level disinfection reliably, every channel, every time, and a complex device's geometry made that harder to guarantee than a simple blood pressure cuff ever would be. The Spaulding Classification exists precisely so that "how sterile does this need to be" is never left to judgment call.

Key exam facts in one table

Fact Detail
Sterilization Kills/removes all microorganisms, including endospores
Disinfection Kills pathogenic organisms; spores may survive
Decontamination Makes an item safe to handle; the least rigorous of the three terms
Spaulding Classification: Critical Enters sterile tissue/bloodstream → requires sterilization
Spaulding Classification: Semicritical Contacts mucous membranes → requires high-level disinfection
Spaulding Classification: Noncritical Contacts intact skin only → requires low/intermediate-level disinfection
Levels of chemical disinfection Low → intermediate → high, based on which organisms each level kills
Physical sterilization methods Moist heat, dry heat, radiation, filtration, incineration
Chemical sterilants Ethylene oxide, hydrogen peroxide (at sterilant concentration)
CDC statistic >99,000 annual U.S. deaths from hospital-acquired infections
Real-world consequence of misclassification Duodenoscope-linked CRE outbreaks from inadequate high-level disinfection

Where Students Get Confused

  • Confusing high-level disinfection with sterilization. High-level disinfection kills everything except large numbers of bacterial spores. Only sterilization guarantees zero survivors, spores included. This distinction is exactly what separates the Critical and Semicritical Spaulding categories.
  • Using "decontamination" and "disinfection" interchangeably. Decontamination is the broadest, least rigorous term, "safe to handle", with no specific claim about which organisms are dead. Disinfection specifically means pathogens are killed, even if some other organisms (including spores) survive.
  • Assuming Spaulding classification depends on a device's physical properties. It depends entirely on what tissue the device contacts in use, not on whether it's rigid, flexible, disposable, or reusable.
  • Assuming any hospital-grade method works for any device. A method appropriate for a noncritical item (like a stethoscope) is not automatically adequate for a semicritical or critical one, exactly the gap that caused real duodenoscope-related outbreaks.

References

  1. Forbes, B. A., Sahm, D. F., & Weissfeld, A. S. Bailey & Scott's Diagnostic Microbiology (11th ed.).
  2. Review of Medical Microbiology and Immunology (13th ed.). Lange Medical Books.
  3. Mohapatra, S. (2017). Sterilization and Disinfection. In Essentials of Neuroanesthesia (pp. 929–944). https://doi.org/10.1016/B978-0-12-805299-0.00059-2
  4. Madigan, M., Martinko, J., Stahl, D., & Clark, D. (2018). Brock Biology of Microorganisms (15th ed.). Pearson Education.
  1. Centers for Disease Control and Prevention. (2008). Guideline for Disinfection and Sterilization in Healthcare Facilities. https://www.cdc.gov/hicpac/pdf/guidelines/Disinfection_Nov_2008.pdf
FAQ

Frequently Asked Questions

What is the difference between sterilization, disinfection, and decontamination?

Sterilization kills or removes all microorganisms, including spores. Disinfection kills pathogenic organisms but may leave spores viable. Decontamination simply makes an item safe to handle, without a specific claim about what's been killed.

What is the Spaulding Classification?

A framework that sorts medical devices into three categories, critical, semicritical, and noncritical, based on the infection risk of how they're used, determining the minimum required level of sterilization or disinfection.

What processing does a critical medical device require?

Full sterilization. Critical devices enter normally sterile tissue, the vascular system, or the bloodstream, so all microbial life, including endospores, must be destroyed.

What processing does a semicritical device like an endoscope require?

At minimum, high-level disinfection, since these devices contact mucous membranes without penetrating sterile tissue.

What are the levels of chemical disinfection?

Low-level disinfectants handle most vegetative bacteria and some fungi/viruses; intermediate-level disinfectants also kill mycobacteria; high-level disinfectants kill everything except large numbers of bacterial spores.

Why did duodenoscope reprocessing failures cause real hospital outbreaks?

Endoscopes are semicritical devices with complex internal channels that are difficult to fully clean. When high-level disinfection wasn't reliably achieved throughout every channel, resistant organisms like CRE survived and were transmitted between patients.

How many deaths occur annually from hospital-acquired infections in the U.S.?

According to the CDC, more than 99,000 deaths occur each year in the United States due to hospital-acquired infections.

What are the main physical methods of sterilization?

Moist heat (autoclaving), dry heat, radiation, filtration, and incineration.

What is the difference between sterilization and disinfection?

Sterilization destroys ALL microorganisms including endospores — used for items entering sterile body tissues. Disinfection destroys most pathogens but not necessarily endospores — used for surfaces and semi-critical devices. A sterilized item is guaranteed free of all life; a disinfected item is free of most pathogens but may harbor resistant spores.

Which microorganisms are most resistant to disinfectants?

Most to least resistant: Prions > bacterial endospores (Bacillus, Clostridium) > mycobacteria > non-enveloped viruses (Poliovirus, Norovirus) > fungi > gram-negative vegetative bacteria > gram-positive vegetative bacteria > enveloped viruses (HIV, HBV, Influenza, SARS-CoV-2). Enveloped viruses are killed even by soap and water.

Why is glutaraldehyde used for endoscope disinfection?

Flexible endoscopes cannot be autoclaved (heat damages optics and electronics). 2% glutaraldehyde achieves high-level disinfection in 20 minutes and sterilization in 10 hours at room temperature, without corroding endoscope materials. However it is toxic — requires ventilation and PPE. OPA and accelerated hydrogen peroxide are safer alternatives.

What concentration of bleach is used for different purposes?

General surfaces: 0.1% (1,000 ppm) — dilute 1:50. Blood/body fluid spill decontamination: 0.5% (5,000 ppm) — dilute 1:10. Prepare fresh daily — sodium hypochlorite degrades rapidly after dilution. Always clean surfaces with water before applying bleach — organic matter inactivates it.

What is the difference between an antiseptic and a disinfectant?

Same chemical, different application and concentration. Antiseptics are formulated safe for living tissue (skin, mucous membranes) — typically lower concentrations. Disinfectants are for inanimate surfaces — often higher concentrations toxic to living cells. Example: 3% H2O2 = antiseptic for wounds; 6% H2O2 = high-level disinfectant for endoscopes.

Why can't ethylene oxide be used for all medical devices?

EtO is toxic, flammable, and carcinogenic — requires specialized ventilation equipment. Sterilized items need 8–12 hours aeration to remove toxic residues before use. The process takes 4–16 hours total and is expensive. Used only when no other method is suitable — primarily heat-sensitive devices like flexible endoscopes, electronics, and certain plastics.

Why is 70% alcohol more effective than 100% alcohol as a disinfectant?

Pure alcohol dehydrates the cell wall too rapidly, causing surface protein coagulation that forms a protective coat preventing penetration. 70% alcohol dehydrates more slowly, allowing penetration through the membrane to denature intracellular proteins throughout the cell. The water component is essential. Effective range: 60–90% concentration.
Acharya Tankeshwar
About Author
Acharya Tankeshwar

Tankeshwar Acharya, MSc (Medical Microbiology)

Tankeshwar Acharya is an Assistant Professor in the Department of Microbiology at Patan Academy of Health Sciences (PAHS), Nepal, where he has been teaching and practicing clinical microbiology for over 14 years. He is the founder of Microbe Online, one of the leading free microbiology education resources on the web, covering bacteriology, mycology, parasitology, immunology, and clinical laboratory diagnostics written from direct experience in both the classroom and the diagnostic laboratory.